Alnor EBT730 Micromanometer Air Flow Meter with Pitot Tube

Special Price $1,818.15 Regular Price $1,955.00
Mitchell Part Number
TSI-EBT730
Manufacturer
Availability
Typically in stock, Call if Urgent
Quick overview
Alnor EBT730 Micromanometer Air Flow Meter with Pitot Tube

Alnor EBT730 Micromanometer Air Flow Meter with Pitot Tube

The Alnor Micromanometer EBT730 is one of the most advanced, versatile, and easy to use micromanometers on the market today. The EBT730 features an auto-zeroing pressure sensor that increases measurement resolution and accuracy along with an intuitive menu structure for ease of operation.

Features and Benefits

  • Accurately measures pressure, velocity and flow to help you meet industry standards
  • Auto-zeroing pressure sensor reduces user-steps and saves time
  • Automatic density correction increases reading accuracy
  • Intuitive menu structure allows for ease of use and setup
  • Integrated Log-Tchebycheff duct traverse application simplifies calculations
  • Bluetooth communications for transferring data or remote polling
  • Includes downloading software with USB cable
  • Accommodates optional pitot, air flow (straight pitot), temperature/relative humidity, velocity matrix, or thermoanemometer probes for use in multiple applications
  • Large graphic display with backlight offers easy-to-use interface
    • Displays up to five measurements simultaneously
    • On-screen messages and instructions
    • Programmed for multiple languages

Applications

  • Energy efficiency studies
  • Fume hood and biological safety cabinet face velocity spot checking
  • HVAC system testing, adjusting and balancing
  • Building and system commissioning
  • Cleanroom testing
More Information
Weight 5.000000
Product Includes Manometer, carrying case, rechargeable NiMH batteries, AC adapter, 18 in.Pitot probe, (2) Static Pressure probes, 16 ft Neoprene tubing, software USB cable, NIST certificate and Manual
Specifications

Velocity Range

Pitot probes 25 to 15,500 ft/min (0.125 to 78 m/s)

Accuracy ±3% of reading ±7 ft/min (±0.04 m/s) at velocities >50 ft/min (>0.25 m/s)
Units ft/min, m/s
Resolution 1 ft/min (0.01 m/s)

Pressure

Differential pressure ±15 in. H2O (±3735 Pa);150 in. H2O (37.5 kPa), maximum safe operating pressure
Absolute pressure 15 to 40 in. Hg (356 to 1016 mm Hg)
Accuracy ±2% of reading ±0.001 in. H2O (±0.25 Pa) static and differential; ±2% of reading absolute
Units in. H2O, in. Hg, Pa, hPa, kPa, mm Hg, cm Hg, mm H2O, cm H2O,
Resolution 0.00001 in. H2O (0.001 Pa) static and differential; 0.01 in. Hg (1 mm Hg) absolute

RH (optional Probe)

Range 0 to 95% RH temperature/RH probe
Accuracy ±3% RH
Resolution 0.1% RH

Temperature (optional Probe)

Sensor in base 40 to 140°F (4.4 to 60°C)
Temperature probe -40 to 250°F (-40 to 121°C) - (Optional)
Temperature/RH probe 14 to 140°F (-10 to 60°C) - (Optional)
Accuracy ±0.5°F (±0.3°C) from 32 to 160°F (0 to 71°C)
Units °F, °C
Resolution 0.1°F (0.1°C)

Instrument Temperature Range

Operating 40 to 140°F (4.4 to 60°C)
Storage -4 to 160°F (-20 to 71°C)


Statistics min, max, average up to 1000 readings
Data Storage 26,500 readings, time and date stamped

size 2’ x 2’ capture hood assembly, additional sizes available
Power Supply Four “AA” batteries or ac adapter
Manufacturer TSI
MPN EBT730
Manufacturer Warranty 1 year